The SAW-96 Automated Nucleic Acid Extraction Station (hereinafter referred to as “SAW-96”), produced by Jiangsu Bioperfectus Technologies Co., Ltd., is an in vitro diagnostic medical device that integrates advanced mechanical, electronic, and software technologies. The SAW-96 combines in a single equipment the functions of sample loading, nucleic acid purification, and PCR preparation. The typical run time is less than 70 minutes for 1–96 samples. Based on experienced magnetic separation technology, the SAW-96 offers high-throughput automation combined with high performance, promoting greater laboratory productivity to meet variable demands. With contactless dispensing of samples, reagents, and PCR reagents, the system is safe and simple for operators, while saving time and costs for the laboratory.
The equipment is intended for automated extraction workflows. It enables automatic operations from primary sample tubes to nucleic acid extraction and PCR preparation.
“Walk-away” station: a single equipment from primary sample tube to PCR-ready tube, including sample loading, nucleic acid purification, and PCR preparation.
High efficiency: processing 1–96 samples with the entire process completed in 70 minutes.
User-friendly design: visual interface with functions displayed in real-time in the software.
Better contamination control: UV light, HEPA filter, and electrostatic filter help prevent aerosols in the internal environment. Scientific zoning, leak-proof design pipetting device, and filter tips increase the level of cross-contamination prevention.
Quality commitment: 18-month warranty.
| Dimensions (W × D × H) |
1.130 × 680 × 690 mm |
| Weight | 110,8kg |
| Sample capacity 1-96 |
|
| Cooling control 4–15 °C for reagent station and 4–8 °C for PCR preparation station. |
|
| Contamination control Nucleic acid extraction and solution dispensing operations are strictly performed in separate areas. An integrated medical purifier consists of two timer-controlled UV disinfection modules, two HEPA filters, and a high-efficiency, high-voltage electrostatic filter for particle capture, with four fans. | |
| Voltage | 100-240 V AC |
| Frequency | 50-60 Hz |
| Rated power 1210 VA |
|
|
Sampling Function |
|
| Compatible sample tubes |
Sample tubes and centrifuge tubes, such as 1.5 mL, 5 mL tubes and tubes with 16 mm diameter. |
| Sample types |
Plasma, serum, whole blood, swab, urine, etc. |
| Extraction Function | |
| Magnetic bead recovery efficiency ≥ 98% |
|
| Processing volume | 20-1,000 μL |
| Mixing mode Oscillatory mixing with multiple modes and speeds (>20 levels) |
|
| Compatible kit Bioperfectus magnetic bead-based nucleic acid extraction kits |
|
| Heating temperature | Lysis heating: room temperature to 120°C |
| Elution heating: room temperature to 120°C | |
|
Pipetting Function |
|
| Sample loading channels | |
| Reagent tubes 1.5 mL, 2 mL, 5 mL EP tubes; 0.2 mL PCR tubes |
|
| Piston pump dispensing volume | 1-250 μL |
| Positioning accuracy ±0.1 mm in X-Y-Z axes |
NMPA, CE, FDA Listed